Level measurement (FF configuration program) Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus
22 Endress+Hauser
4 Select the measuring mode if necessary:
• Select the "No linearization" option by means of the
Linearization/LINEARIZATION parameter.
• By means of the Primary Value Type/
PRIMARY_VALUE_TYPE parameter, select the
"Volume" option.
P01-xxxxxxxx-05-xx-xx-xx-029
P01-xxxxxxxx-05-xx-xx-xx-032
Fig. 8: Calibration with reference pressure –
wet calibration
1 See Table, Step 10.
2 See Table, Step 12.
3 See Table, Step 13.
4 See Table, Step 14
5 See Table, Step 15.
5 Select the "Level easy height" option by means of the
Level Selection/LEVEL_SELECTION parameter.
6 Select a volume unit, such as "l" in this example, by
means of the Output Unit Level Easy/
OUTPUT_UNIT_EASY parameter.
7 Select a height unit by means of the Height Unit Level
Easy/HEIGHT_UNIT_EASY parameter, here "m" for
example.
8 By means of the Calibration Mode Level Easy/
CALIBRATION_MODE_EASY parameter, select the
"Dry" option.
9 Select a density unit via the Density Unit Level Easy/
DENSITY_UNIT_EASY parameter, here "kg/dm
3
" for
example.
10 By means of the Adjust Density Level Easy/
ADJUST_DENSITY_EASY parameter, enter a density,
here "1" "kg/dm³" for example.
11 By means of the Empty Calib. Level Easy/
EMPTY_CALIBRATION_EASY parameter, enter a
volume, here 0 l for example.
12 By means of the Empty Height Level Easy/
EMPTY_HEIGHT_EASY parameter, enter a volume,
here 0.5 m (1.6 ft) for example.
13 By means of the Full Calib. Level Easy/
FULL_CALIBRATION_EASY parameter, enter a volume,
here 1000 l (264 US gal) for example.
14 By means of the Full Height Level Easy/
FULL_HEIGHT_EASY parameter, enter a volume, here
4.5 m (15 ft) for example.
15 Set the Pressure Transducer Block to the "Auto" block
mode.
Description
4.5
0.5
h
[m]
49 441
p
[mbar]
h =
p
r · g
r = 1
g
cm
3
➀
1000
0
V
[l]
0.5 4.5
➁
➂
h =
p
ρ · g
h
[m]
➃
➄