Gammapilot FMG50
Endress+Hauser 11
A0018248
A One-point calibration
B Multiple point calibration
D Pipe diameter or radiated length
I Pulse rate (pulses per second, cnt/s)
F
S
Attenuation factor
ρ Density
µ Absorption coefficient
Concentration measurement
The Gammapilot FMG50 determines the concentration indirectly via a density measurement. A
linearization table consisting of up to 32 "density - concentration" value pairs can be entered for this
calculation. The solids content of liquids, for example, can be measured in this way (percentage of
volume or weight).
A0018249
I Pulse rate (pulses per second, cnt/s)
ρ Density
C Concentration
Concentration measurement with radiating media
The Gammapilot FMG50 calculates the concentration of the medium from the intensity of the
radiation that is emitted by the medium itself.
A source container and radiation source are not needed for the measurement