Gammapilot FMG50
16 Endress+Hauser
1
A0036646
1 Gammapilot FMG50
Density measurement
In the case of density measurement, the density measuring range is defined by the minimum and
maximum density of the measured medium.
1
A0036645
1 Gammapilot FMG50
Interface measurement
In the case of interface measurement, the measuring range is determined by the possible position of
an interface. Here, the 0% position is at the lowest possible point where the interface is to be
monitored, while the 100% position is at the highest point.
Concentration measurement with source container and radiation source
In the case of concentration measurement, the measuring range is defined by the minimum and
maximum concentration of the measured medium.
Concentration measurement with radiating media
In the case of radiating media, the measuring range is defined by the minimum and maximum
concentration of the medium.
Measurement of mass flow
For the FMG50, mass flow measurement constitutes a density measurement.
The density measuring range is defined by the minimum and maximum density of the measured
medium.
Conditions/prerequisites for applications in safety-related operation
See the Functional Safety Manual
"Slave" operating mode: in this measuring mode, the measured pulse rate is output for further
processing in a connected controller.
This mode is not permitted for "functional safety"