Gammapilot FMG50
22 Endress+Hauser
Connection of the HART device and RIA15 with backlighting
A0019568
6 Block diagram of HART device with RIA15 process indicator with light
1 Device with HART communication
2 Power supply
3 HART resistor
FMG50, RIA15 with installed HART communication resistor module
The HART communication module for installation in the RIA15 can be ordered together with
the device.
Product structure, feature 620 "Accessory enclosed":
Option PI "HART communication resistor for RIA15"
The voltage drop to be taken into account is:
Max. 7 V
Alternatively available as an accessory, for details see Technical Information TI01043K and
Operating Instructions BA01170K
Connection of the HART communication resistor module, RIA15 without backlighting
A0020839
7 Block diagram of HART device, RIA15 without light, HART communication resistor module
1 HART communication resistor module
2 Device with HART communication
3 Power supply
Connection of the HART communication resistor module, RIA15 with backlighting
A0020840
8 Block diagram of HART device, RIA15 with light, HART communication resistor module
1 HART communication resistor module
2 Device with HART communication
3 Power supply