EasyManua.ls Logo

Endress+Hauser iTEMP HART TMT 182 - Installing in a Sensor Connection Head; Installation in a Field Housing

Endress+Hauser iTEMP HART TMT 182
86 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Temperature head transmitter 3 Installation
Endress+Hauser 37
3.2 Installation
Installation For installation proceed as follows:
Installing in a sensor connection head to DIN 43 729 Form B (s. fig. 3-1, left side)
Feed the sensor inset cables (Pos. 5) through the central hole in the head transmitter
(Pos. 4).
Place the installation springs (Pos. 3) onto the screws (Pos. 2).
Feed the installation screws (Pos. 2) through the holes in the head transmitter and the
holes in the sensor inset (Pos. 5). Fix both screws using the circlips (Pos. 6).
Position the head transmitter in the connection head in such a way so that the current
output terminals (terminal 1 and 2) are towards the cable entry gland (Pos. 7).
Fix the head transmitter (Pos. 4) and sensor inset (Pos. 5) into the connection head.
Installation in a field housing (s. fig. 3-1, right side)
Feed the installation screws (Pos. 2) with installation springs (Pos. 3) through the holes
in the head transmitter (Pos. 4). Fix these using the circlips (Pos. 5).
Screw the head transmitter into the field housing using a screwdriver.
"
Attention!
In order to avoid damaging the head transmitter, do not over-tighten the installation
screws.
Fig. 3-1: Installation of head transmitter into a Form B sensor connection head (left side) and a field housing
(right side).
R-TMT187ZX-17-06-06-en-001.eps
R-TMT187ZX-17-06-06-en-002.eps