Commissioning Levelflex M FMP40 FOUNDATION Fieldbus
50 Endress+Hauser
6 Commissioning
This chapter consists of the following sections:
"Function check", ä 50
"Unlocking the device", ä 50
"Resetting the device", ä 52
"Basic setup", ä 54
"Commissioning by the display and operating module VU331", ä 56
"Blocking distance", ä 66
"Envelope curve", ä 68
"Basic setup with the Endress+Hauser operating program", ä 72
"Commissioning with a FOUNDATION Fieldbus configuration tool", ä 78
"Commissioning with the handheld terminal Field Communicator 375, 475", ä 82
6.1 Function check
Make sure that all final checks have been completed before you start up your measuring point:
Checklist "Post installation check", ä 33.
Checklist "Post connection check", ä 38.
6.2 Unlocking the device
Before commissioning, make sure that the device is not locked against parameter changes.
On delivery, the device is unlocked. In other cases, however, it may have been locked in one of the
following ways:
6.2.1 DIP switch (under the housing cover)
Locking and unlocking
L00-FMxxxxxx-19-00-00-en-002
WP = on: parametrization locked
WP = off: parametrization unlocked
SIM = on: simulation possible in Analog Input Block by configuration tool
SIM = off: simulation not possible in Analog Input Block by configuration tool
Parameters affected
Locking by the DIP switch affects all parameters.
E
N
D
R
E
S
S
+
H
A
U
S
E
R
M
I
C
R
O
P
I
L
O
T
I
I
ENDRESS+HAUSER
MICROPILOT II
I
P
6
5
IP 65
O
r
d
e
r
C
o
d
e
:
S
e
r
.
-
N
o
.
:
Orde
r
C
od
e:
S
e
r.-No.:
M
e
s
s
b
e
r
e
i
c
h
M
e
a
s
u
r
i
n
g
r
a
n
g
e
Messbere
i
ch
Measuri
ng
range
U
1
6
.
.
.
3
6
V
D
C
4
.
.
.
2
0
m
A
U
1
6
.
.
.
36
V
D
C
4...
2
0
mA
m
a
x
.
2
0
m
m
ax.20 m
Made in Germany Maulburg
Made in Germany Maulburg
T>
7
0
°
C
:
A
t
>
8
5
°
C
T>7
0
°
C :
A
t >85°C
on
off
SIM WP
SIM WP
off
on
simulation
write protection
default settings:
write protection OFF
simulation OFF