EasyManua.ls Logo

Endress+Hauser Levelflex M FMP40 - Basic setup

Endress+Hauser Levelflex M FMP40
128 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Levelflex M FMP40 FOUNDATION Fieldbus
54 Endress+Hauser
6.4 Basic setup
L00-FMP4xxxx-19-00-00-en-031
SD
UB
F
L
D
E
LN
flange:
reference point of
measurement
tank shape
Standard
measuring cond.
Unknown
measuring cond.
Standard
mapping
safety settings
linearisation
extended calibr.
...
empty calibr.
full calibr.
(description see BA00245F/00/EN)
basic setup (standard)
option
E = empty calibr. (= zero) LN = probe length
setting in 005 setting in 033
F = full calibr. (= span) UB = upper blocking distance
setting in 006 setting in 059
D = distance (distance flange / product) SD = safety distance
display in 0A5 setting in 015
L = level
display in 0A6

Table of Contents