EasyManua.ls Logo

Endress+Hauser Levelflex M FMP40 - Page 79

Endress+Hauser Levelflex M FMP40
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Levelflex M FMP40 FOUNDATION Fieldbus Commissioning
Endress+Hauser 79
6.9.2 Parametrization of the Resource Block (Start Index: 400)
1. Enter the desired block name (optional).
Factory setting: RESOURCE_XXXXXXXX
2. Opern the Resource Block
3. On delivery, write protection is disabled so that you can access the write parameters via
FOUNDATION Fieldbus. Check this status by the parameter WRITE_LOCK:
Write protectin activated: WRITE_LOCK = LOCKED
Write protection deactivated: WRITE_LOCK = NOT LOCKED
Deactivate the write protection if necessary, ä 50.
4. Set the operating mode to AUTO in the parameter group MODE_BLK (parameter TARGET).
6.9.3 Parametrization of the Sensor Block (Start Index: 2000)
1. Enter the desired block name (optional)
Factory setting: SENSOR_XXXXXXXX
2. Open the Sensor Block. The following display appears:
NI-FBUS-FMU4x-en-003
!
Note!
There are two possibilities to edit the parameters of the block:
A parameter from the list can be opened for editing by a double click.
You can open one of the FOUNDATION Fieldbus methods. Each method guides you
automatically through a number of parameters which are required for a specific configuration
task. The following sections describe the parametrization by the "basic setup" method.

Table of Contents

Other manuals for Endress+Hauser Levelflex M FMP40

Related product manuals