Liquistation CSF34 Product description
Endress+Hauser 11
3.2 Equipment architecture
3.2.1 Slot and port assignment
E
85
85
86
86
41
43
42
87
88
97
98
31
31
32
32
SD
Display
1
2
Sensor
0/4 ... 20 mA
Sensor supply
+
+
+
+
1
2
Service
PK
GY
PK
GYGY
1
2
BN
WH
GN
YE
BN
WH
GN
YE
87
88
97
98
+
AB
+
A
B
Alarm
HART
Power
(internal)
24VDC
24V
12,5V
3,3V
A0045689
 3 Slot and port assignment of hardware and presentation on the display
The electronics configuration follows a modular concept:
There are several slots for electronics modules. These are referred to as "slots".
These slots are numbered consecutively in the housing. Slots 0 and 1 are always reserved
for the base module.
In addition there are also inputs and outputs for the control module. These slots are
labeled "S".
Each electronics module has one or more inputs and outputs or relays. Here they are all
collectively known as "ports".
Ports are consecutively numbered per electronics module and are recognized
automatically by the software.
Outputs and relays are named according to their function, e.g. "current output", and are
displayed in ascending order with the slot and port numbers.
Example:
"Current output 2:1" shown on the display means: slot 2 (e.g. AOR module) : port 1
(current output 1 of the AOR module)
Inputs are assigned to measuring channels in the ascending order of "slot:port number"
Example:
"CH1: 1:1" shown on the display means:
Slot 1 (base module) : port 1 (input 1) is channel 1 (CH1).