Diagnostics and troubleshooting Liquistation CSF34
172 Endress+Hauser
11.11 Resetting the device
Menu/Diagnostics
Function Options Info
Device restart Selection
OK
ESC
Restart and keep all the settings
Factory default Selection
OK
ESC
Restart with factory settings Settings that have
not been saved are lost.
11.11.1 Operating time information
The following information is displayed:
Operating hours device:
Displays the total operating hours of the device in days, hours and minutes
Operating hours cooling (only for version with climate control module):
Displays the total operating hours of the compressor in days, hours and minutes
Overfill sensor (for version with vacuum pump):
Number of times a safety switchoff has been caused by LF2
Dosing valve (for version with vacuum pump):
Number of times the dosing valve is actuated; > corresponds to the number of samples
taken
Vacuum pump (for version with vacuum pump):
Displays the pump operating time in hours and minutes
Sample totalizer (for version with peristaltic pump):
Number of all samples taken and sample errors
Pump tube life (for version with peristaltic pump):
Displays how old the tube is in days, hours and minutes
Peristaltic pump (for version with peristaltic pump):
Displays the pump operating time in hours and minutes
This counter must be reset when a tube is replaced.
Filter mats:
Displays the period of usage in days
Operating time photometer:
Displays the hours of operation in hours
Operating time cooling module:
(optional): Displays the period of usage in days.
With Reset, you reset the counter to zero.
11.11.2 Status of inputs/outputs
Path: MEAS/Measurement
The following measured values are listed (read only):
Binary inputs
Current function state: on or off
Alarm relay
Current function state: on or off
Binary outputs
Current function state: on or off
Temperature sensors
Current value is displayed: S:1 (cooling system)
Current outputs
(for version with sensors with the Memosens protocol) Actual current values of the
current outputs