Liquistation CSF34 Mounting
Endress+Hauser 21
Connecting the suction line on version with peristaltic pump
1
2
3
4
A0013705
 12 Connecting the suction line from the side (as-delivered
state)
1 Small thread adapter nut
2 Hose
3 Thread adapter nut
4 Hose gland
1
A0013706
 13 Suction line connected from below
Modification of the suction line from side connection to connection from below
1. Unscrew the thread adapter nut (item 3) and the hose gland (item 4) from the side
panel.
2. Unscrew the small thread adapter nut (item 1) and remove the hose.
3. Fit a hose adapter on the hose.
4. Secure the hose adapter with a worm drive hose clip.
5. Connect the suction line from below as illustrated.
6. Insert the dummy plugs supplied.
5.3 Sampling with a flow assembly
The sample is extracted either directly from the flow assembly which is installed in the
base or from an external flow assembly.
The flow assembly is used for sampling in pressurized systems e.g.:
Tanks positioned at a height
Pressure piping
Conveyance using external pumps
The max. flow rate should be 1000 to 1500 l/h.
1
2
3
A0013127
 14 Connections on flow assembly 71119408
1
2
3
Flow assembly inflow: ¾"
Sampling connection
Flow assembly outflow: 1¼"
The outlet of the flow assembly must be unpressurized (e.g. drain, open channel).