Technical data Liquistation CSF34
202 Endress+Hauser
EtherNet/IP
Log EtherNet/IP
ODVA certification Yes
Device profile Generic device (product type: 0x2B)
Manufacturer ID 0x049E
h
Device type ID 0x109
Polarity Auto-MIDI-X
Connections CIP 12
I/O 6
Explicit message 6
Multicast 3 consumers
Minimum RPI 100 ms (default)
Maximum RPI 10000 ms
System integration EtherNet/IP EDS
Rockwell Add-on-Profile Level 3, Faceplate for Factory
Talk SE
IO data Input (T → O) Device status and diagnostic message with
highest priority
Measured values:
16 AI (analog input) + Status + Unit
8 DI (discrete input) + Status
Output (O → T) Actuating values:
4 A0 (analog output) + status + unit
8 DO (discrete output) + Status
Web server The web server enables full access to the device configuration, measured values, diagnostic
messages, logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G routers
with a user-defined IP address.
TCP port 80
Supported features Remote-controlled device configuration
Save/restore device configuration (via SD card)
Logbook export (file formats: CSV, FDM)
Access to web server via DTM or Internet Explorer
15.4 Power supply
Supply voltage 100 to 120/200 to 240 V AC ±10 %, 50/60 Hz
The device does not have a power switch.
A fuse with a maximum rating of 10 A must be provided by the customer. Observe the
local regulations for installation.
Power consumption Version with vacuum pump: 290 VA
Version with peristaltic pump: 290 VA
Version with sampling assembly: 290 VA
Version with 24V power supply: 240 W
Electrical connection See the "Electrical connection" section ()