Operation Liquistation CSF34
112 Endress+Hauser
Setup program
Function Info
Edit The selected program is displayed and can be edited. Press the "SAVE"
button to save the changes.
Delete The selected program is deleted following a confirmation prompt.
Start The selected program is started immediately. The program can be
canceled or paused by pressing the OFFbutton. If there are differences
between the setup and the selected program, the Program configuration
contains errorsmessage appears, e.g. the bottle configuration in the
program does not match the configuration in the setup. The program is
not started. In this example, the actual bottle configuration must be
checked against the configuration in the setup and the program and
changed accordingly. Only the bottle configuration entered in the setup is
valid for the program to be executed.
Duplicate The selected program is duplicated and saved with an ID.
Cancel Back to the overview.
The Setup programdisplay features the ESC, MAN, ? and MODE softkeys.
The Program enableddisplay features the ESC, STAT and MODE softkeys.
Setup program
Function Info
ESC Back to the overview. Any program currently running is canceled.
MAN Manual sampling can be configured and started here. Any program
currently running is paused. > See "Sampling program/Manual sampling"
section
? A help text is displayed for the item.
STAT For selecting statistics about measured values, sampling and inputs, see
the "Display behavior" section.
MODE If no program is enabled, the device can be switched off here. If a program
is enabled, the following options appear:
Power down sampler:
Following a confirmation prompt, the device is set to the standby mode.
Power continues to be supplied to the device and the LED flashes green.
The display goes dark.
Stop program %0V:
1)
Stops a program currently running following a confirmation prompt. The
overview display appears.
Pause program %0V:
Is selected if maintenance work is pending. The program is paused and
the pause time is entered in the logbook. After pressing the Resume
program button, the current program is continued.
1) "%0V" here stands for text that depends on the context. This text is generated automatically by the
software and inserted in place of %0V. In the simplest scenario, this could be the name of the measuring
channel, for example.
10.4 Inputs
As standard, Liquistation CSF34 has:
Two binary inputs
Two current inputs
Galvanically isolated from one another