Liquistation CSF34 Technical data
Endress+Hauser 205
Process pressure range • Unpressurized, open channel (unpressurized sampling)
• Max. 0.8 bar piping (only with shutoff/inlet valve)
Sampling assembly:
Max. 6 bar
Medium properties Vacuum pump
Capacitance level measurement used for:
• Sample media has to be free of abrasive substances.
• Media that tend to create a lot of foam or contain fats and grease
• Media with a conductivity < 30 µS/cm
Peristaltic pump
Sample media has to be free of abrasive substances.
Pay attention to the material compatibility of the wetted parts.
Process connection • Vacuum pump:
Intake hose ID 10 mm (3/8 in), 13 mm (1/2 in), 16 mm (5/8 in) or 19 mm (3/4 in)
• Peristaltic pump:
Intake hose ID 10 mm (3/8 in)
• Sampling assembly:
• Flange DN50, PP
• Triclamp DN50, DIN 32676
15.8 Mechanical construction
Dimensions See the "Installation" section → 15
Weight
Sampler version Weight
Plastic version with refrigeration 101 kg (223 lbs)
Materials Plastic polystyrene VO can change color when exposed to direct sunlight. For outdoor
use without a weather protection cover, the use of Plastic ASA+PC V0 is
recommended. The functionality is not affected by the discoloration.
Non-wetted parts
Cabinet housing Plastic ASA+PC V0
For industrial wastewater treatment plants with an aggressive
atmosphere
Sample compartment inner lining Plastic PP
Window Safety glass, coated
Insulation Plastic EPS "Neopor®"
Choose process seal depending on the application. Viton is recommended for standard
applications involving watery samples.
Vacuum pump only
Pneumatic hoses Silicone
Air Manager housing PC