Liquistation CSF34 Mounting
Endress+Hauser 19
Requirements for the sampling point:
Do not connect the suction line to pressurized systems.
Use the suction filter to impede coarse and abrasive solids and solids which can cause
clogging.
Immerse the suction line in the direction of flow.
Take the sample at a representative point (turbulent flow, not directly at the bottom of
the channel).
Useful sampling accessories
Suction filter:
Impedes coarser solids and solids which can cause clogging.
5.2 Setting up the device
5.2.1 Connecting the suction line at the side on version with pump
1. When setting up the device, take the installation conditions into account.
2. Route the suction line from the sampling point to the device.
3. Fit a hose adapter on the hose.
4. Secure the hose adapter with a worm drive hose clip.
5. Screw the suction line onto the device's hose connection.
5.2.2 Connecting the suction line from below on version with pump
If the suction line is connected from below, the suction line is routed upwards behind the
rear panel of the sample compartment.
1. Remove the rear panel of the dosing compartment and sample compartment
beforehand.
2. Remove the drain plug from the hose gland located at the back of the device base.
3. Guide the suction line upwards and through the opening towards the front, as
illustrated.