Liquistation CSF34 Technical data
Endress+Hauser 201
15.3 Protocol-specific data
HART
Manufacturer ID 11
h
Device type 119D
h
Device revision 001
h
Device description files (DD/DTM) www.endress.com/hart
Device Integration Manager DIM
Device variables
Supported features PDM DD, AMS DD, DTM,
PROFIBUS DP
Manufacturer ID 11
h
Device type 155C
h
Profile version 3.02
Device database files (GSD files) www.endress.com/profibus
Device Integration Manager DIM
Output variables
Supported features 1 MSCY0 connection (cyclical communication, master class 1
to slave)
1 MSAC1 connection (acyclical communication, master class 1
to slave)
2 MSAC2 connections (acyclical communication, master class
2 to slave)
Addressing using DIL switches or software
GSD, PDM DD, DTM
Modbus RS485
Protocol RTU/ASCII
Function codes 03, 04, 06, 08, 16, 23
Broadcast support for function codes 06, 16, 23
Output data 16 measured values (value, unit, status), 8 digital values (value,
status)
Input data 4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Supported features Address can be configured using switch or software
Modbus TCP
TCP port 502
TCP connections 3
Protocol TCP
Function codes 03, 04, 06, 08, 16, 23
Broadcast support for function codes 06, 16, 23
Output data 16 measured values (value, unit, status), 8 digital values (value,
status)
Input data 4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Supported features Address can be configured using DHCP or software