Liquistation CSF34 Operation
Endress+Hauser 117
Menu/Setup/Inputs
Function Options Info
Reset totalizer Selection
Manual
Automatic
At program start
Factory setting
Manual
Manual
Reset the counter manually.
Automatic
The counter is reset automatically at
intervals.
At program start
The counter is reset at program start.
Flow rate - - - The current flow rate is displayed.
If you select Reset totalizer Manual
Reset totalized flow Action The totalized flow currently calculated is set
to zero when the counter is reset.
If you select Reset totalizer Automatic
Interval Selection
Daily
Weekly
Monthly
Factory setting
Daily
Daily
If a daily interval is selected, set the Time in
the following menu item.
Weekly
If a weekly interval is selected, set the Day
of week and the Time in the following
menu items.
Monthly
If a monthly interval is selected, set the Day
of month and the Time in the following
menu items.
If you select Input variable Parameter
Meas. value format Factory setting
#.#
Specify the number of decimal places.
Parameter name User-defined text Assign a name.
Unit of measure User-defined text Enter the engineering unit.
Lower range value -20 to 10000
Factory setting
0
Enter a value for the start of the measuring
range. 0/4 mA is assigned to this value as
per your specifications.
-20 to 10000
Factory setting
10
Enter a value for the end of the measuring
range. 20 mA is assigned to this value as per
your specifications.
Upper range value 0 to 60 s
Factory setting
0 s
The damping causes a floating average curve
of the measured values over the time
specified.
If you select Input variable Current
Meas. value format Factory setting
#.#
Specify the number of decimal places.
Damping 0 to 60 s
Factory setting
0 s
The damping causes a floating average curve
of the measured values over the time
specified.
10.5 Outputs
10.5.1 Binary outputs
The basic version of the device always has two binary outputs.