Operation Liquistation CSF34
118 Endress+Hauser
Possible application --> For outputting a manipulated variable to connected actuators
The binary output must be assigned in the program or subprogram before it can be
activated.
Menu/Setup/Outputs
Function Options Info
OutputBinary
Function Selection
• Off
• Event
• Limit value
• Diagnostics message
• Cleaning
(Only for version with
sensors with Memosens
protocol)
Factory setting
Off
The following functions depend on the
option selected. Function = "Off" switches off
the function of the binary output and means
no further settings are required.
For Cleaning:
The outputs can only switch 100mA. A
relay must be added to control valves
or motors.
If you select Function Event
Signal slope Selection
• Low-High
• High-Low
Factory setting
Low-High
Select the level change of the signal