Liquistation CSF34 Operation
Endress+Hauser 119
Menu/Setup/Outputs
Function Options Info
Event Selection
• Program enabled
• End of program
• Sampling start
• End of sampling
• End of mult. samp.
• Dosing
• Sampling cycle
• Bottle change
• External stop
• No sample
• Sub program enabled
• Sub prog. activ.
• Sub prog. deactiv.
Factory setting
Sampling cycle
Program enabled
A permanent signal is switched when the
sampling program starts.
End of program
A pulse or permanent signal is switched
when the sampling program ends.
Sampling start
A pulse is switched when a sample is taken.
End of sampling
A pulse is switched when sampling has
ended.
End of mult. samp.
A pulse is switched when sampling of the
last multiple bottle has ended.
Dosing
A pulse is switched at the start of dosing.
Sampling cycle
The output signal is switched for the
duration of the sampling cycle.
Bottle change
A pulse is switched when a bottle is
changed.
External stop
A pulse is switched when an external stop is
performed.
No sample
The output signal is switched if no sample
was taken.
Sub program enabled
The output signal is switched if this
subprogram is active.
Sub prog. activ.
The output signal is switched when the
subprogram starts.
Sub prog. deactiv.
The output signal is switched when the
subprogram ends.
If you select Function Limit value
Signal slope Selection
• Low-High
• High-Low
Factory setting
Low-High
Preselect the level change of the signal.
Source of data Selection
• None
• Limit switch 1-8
Factory setting
None
Select the limit switch via which the status
of the relay is to be output. The limit
switches are configured in the "Setup/
Additional functions/Limit switch" menu.
If you select Function Diagnostics message
Signal slope Selection
• Low-High
• High-Low
Factory setting
Low-High
Preselect the level change of the signal.