Operation Liquistation CSF34
120 Endress+Hauser
Menu/Setup/Outputs
Function Options Info
Operating mode Selection
• as assigned
• Namur M
• Namur S
• Namur C
• Namur F
Factory setting
as assigned
as assigned
If this option is selected, the diagnostics
messages which you have individually
assigned to the binary output are output via
the binary output.
Namur M to F
If you decided to use one of the Namur
classes, all the messages that are assigned to
the individual class are output via the binary
output. You can also change the Namur class
assignment for every diagnostic message.
(Menu/Setup/General settings/
Diagnostics/Device behavior or Menu/
Setup/Inputs/../Diagnostics settings/Diag.
behavior)
Attributed diagnostic
messages
Read-only list of diagnostic
messages
All the messages assigned to the relay
output appear on the display. You do not
have the option of editing the information
here.
If you select Function Cleaning (only for version with sensors with the Memosens protocol)
Signal slope Selection
• Low-High
• High-Low
Factory setting
Low-High
Preselect the level change of the signal.
Assignment Selection
• None
• Cleaning 1-4
Factory setting
None
Use this function to choose the cleaning
instance which should be started when the
binary output is active.
10.5.2 Current outputs
Two current outputs are optionally available on the base module‐E.
Setting the current output range
‣
Menu/Setup/General settings: 0..20 mA or 4..20 mA.
Possible applications
• For outputting a measured value to a process control system or an external recorder
• For outputting a manipulated variable to connected actuators
The current output curve is always linear.
Menu/Setup/Outputs/Current output x:y
1)
Function Options Info
Current output Selection
• Off
• On
Factory setting
Off
Use this function to activate or deactivate a
variable being output at the current output
Source of data Selection
• None
• Connected inputs
Factory setting
None
The sources of data on offer depend on your
device version.