Liquistation CSF34 Diagnostics and troubleshooting
Endress+Hauser 151
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
310 Temperature sensor F On On Temperature sensor PT1 in the climate control
module for sample compartment measurement is
defective
No temperature regulation possible for the
sample compartment
Unable to cancel the sampling program
Contact Endress+Hauser Service.
311 Temperature sensor F On On Temperature sensor PT2 in the sample
compartment is defective
No sample temperature measurement possible
An in-progress sampling program cannot be
canceled
Replace the sensor.
312 Temperature sensor F On On Temperature sensor PT3 for ambient temperature
measurement is defective
Winter operation regulation not possible
Sampling and distribution arm blocked to
protect against freezing
1. Deactivate winter operation under Setup/
Inputs/Temperature S:3/Winter
operation.
2. Replace the sensor.
313 Safety sensor M On On Safety switch LF2 for sample sensor active
Contact electrodes for sample detection are
fouled
The sample continues to be taken
1. Clean sample detection sensor LF1 in the
dosing glass.
2. Contact Endress+Hauser Service.
314 No sample flow F On On A vacuum cannot be generated in the peristaltic
pump.
1. Check the pump tube for leaks.
2. Immerse the suction line in the medium.
315 Refrigeration F On On Sample compartment target temperature not
reached
Cooling regulation not possible
1. Check the sample compartment door.
2. Perform the module test under Menu/
Diagnostics/System test/Cooling system/
Check cooling.
3. Contact Endress+Hauser Service.
316 Heating F On On Sample compartment target temperature not
reached
Heating regulation not possible
1. Check the sample compartment door.
2. Perform the module test under Menu/
Diagnostics/System test/Cooling system/
Check heating.
3. Contact Endress+Hauser Service.
317 Liquidsensor M On On Sensor LF1 for sample detection fouled
Five samples still possible
Clean sensor LF1 in the dosing glass.
318 Liquidsensor F On On Sensor LF1 for sample detection defective
No sampling possible
Contact Endress+Hauser Service.