Operation Liquistation CSF34
62 Endress+Hauser
Menu/Setup/General settings/Date/Time
Function Options Info
Extended setup
Date format Selection
• DD.MM.YYYY
• YYYY-MM-DD
• MM-DD-YYYY
Factory setting
DD.MM.YYYY
‣
Select a date format.
Time format Selection
• hh:mm am (12h)
• hh:mm (24h)
• hh:mm:ss (24h)
Factory setting
hh:mm:ss (24h)
‣
Choose between 12-hour display or 24-hour
display. Seconds can also be displayed with
the latter version.
Time zone Selection
• None
• Choice of 35 time zones
Factory setting
None
None = Greenwich Mean Time (London).
DST Selection
• Off
• Europe
• USA
• Manual
Factory setting
Off
The controller adapts the summertime/normal
time changeover automatically if you choose
European or American daylight saving time.
Manual means that you can specify the start and
end of daylight saving time yourself. Here, two
additional submenus are displayed in which you
specify the changeover date and time.
10.2.3 Hold settings
Menu/Setup/General settings/Hold settings
Function Options Info
Settings automatic Hold
Hold release time 0...600 s
Factory setting
0 s
The hold is maintained for the duration of the
delay time when you switch to the measuring
mode.
Setup menu Selection
• Disabled
• Enabled
Factory setting
Disabled
Diagnostics menu Decide whether a hold should be output at the
current output when the particular menu is
opened.
Calibration active Factory setting
Enabled
If a device-specific hold is enabled, any cleaning that was previously started is
stopped. You can only start a manual cleaning if a hold is active. The hold has no
influence on the sampling.
10.2.4 Logbooks
Logbooks record the following events:
• Calibration/adjustment events
• Operator events
• Diagnostic events
• Programming events