Liquistation CSF34 Operation
Endress+Hauser 93
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
If is selected: Sampling volume input Binary input
Sampling volume 10 to 10000 ml
Factory setting
20 ml
Set the sampling volume.
The dosing accuracy and the repeatability of
a sample volume < 20 ml can vary,
depending on the specific application.
If is selected: Sampling volume input Current input
Sampling volume 20mA 10 to 10000 ml
Factory setting
100 ml
Set what sampling volume should be
grabbed at 20 mA.
The dosing accuracy and the repeatability of
a sample volume < 20 ml can vary,
depending on the specific application.
Flow calculation Selection
• Current
• Average flow
Factory setting
Current
Current:
The current flow is converted to the
sampling volume at the time of sampling.
Average flow:
The system calculates the mean between the
last and the current sample and sets the
sampling volume accordingly.
Bottle change mode Selection
• Number of samples
• Time
• External signal
Factory setting
External signal
The bottle can be changed after a specific
number of samples, after a time or by an
external signal.
Having selected Bottle change mode Number of samples
Samples per bottle 1 ... 9999
Factory setting
1
Set the number of samples.
Having selected Bottle change mode Time
Time interval 00-00:02 ... 31-00:00
DD-HH:MM
Factory setting
00‐01:00 DD‐HH:MM
Set the time (days, hours and minutes) after
which the system should change to the next
bottle.
Bottle synchronization Selection
• None
• 1. bottle change time
• 1. Time of change +
bottle number
Factory setting
None
None
The sampling and bottle change times are
not synchronized.
1. bottle change time
Sampling starts with the first bottle. Set the
synchronization time.
1. Time of change + bottle number
Each bottle is assigned to a specific fill time.
Set the synchronization time and the
weekday.
Start condition Selection
• Immediate
• Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable
time.
Having selected Start condition Date/time
Start date 01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Start time 00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.