Micropilot FMR50/FMR51/FMR52/FMR53/FMR54/FMR56/FMR57 FOUNDATION Fieldbus The "Expert" menu
Endress+Hauser 129
3.9.2 Expert → Diagnostics → Event logbook
Filter options (0705)
Write access Maintenance
Navigation Expert → Diagnostics → Event logbook → Filter options
Description Use this function to select the category whose event messages are displayed in the events
list.
The status signals are categorized according to NAMUR NE 107: F = failure, M =
maintenance request, C = function check, S = out of specification
Options • All
• Failure (F)
• Mainenance required (M)
• Function check (C)
• Out of specification (S)
• Information (I)
Factory setting All
Event list
Navigation Expert → Diagnostics → Event logbook → Event list
Description Use this function to display the history of event messages of the category selected in the
Filter options parameter (→ 129). A maximum of 20 event messges are displayed in
chronological order. If the advanced HistoROM function is enabled in the device, the event
list can contain up to 100 entries.
The following symbols indicate whether an event has occurred or has ended:
• : Event has occurred
• : Event has ended
Information on what is causing the message, and remedy measures, can be viewed via
the -button.
User interface • For event messages in category I: information event, event text, "recording event" symbol
and time event occurred
• For event messages in category F, M, C, S (status signal): diagnostic event, event text,
"recording event" symbol and time event occurred