Micropilot FMR50/FMR51/FMR52/FMR53/FMR54/FMR56/FMR57 FOUNDATION Fieldbus The "Expert" menu
Endress+Hauser 43
3.2.3 Expert → Sensor → Medium
Overview
The Medium submenu is used to specify the relevant properties of the measured medium,
especially the dielectric constant (DC).
Parameter Meaning Description
Medium type Options: "Liquid" or "Solid" (→ 43)
Medium group
1)
Rough selection of the DC ("Water based" or
"Others")
(→ 43)
Medium property Finer selection of the DC (→ 44)
Calcualted DC value Displays the calculated DC. (→ 44)
1) Only available for "Medium type" = "Liquid".
Description of parameters
Medium type (1049)
Navigation Expert → Sensor → Medium → Medium type
Write access Service
Description Indicates the medium type
Display options • Liquid
• Solid
Factory setting • FMR50: Liquid
• FMR51: Liquid
• FMR52: Liquid
• FMR53: Liquid
• FMR54: Liquid
• FMR56: Solid
• FMR57: Solid
Medium group (1208)
Navigation Expert → Sensor → Medium → Medium group
Write access Maintenance
Prerequisite Medium type = Liquid
Description Defines the medium group of the measured product.
Options • Water based (DC ≥ 4)
• Others (DC > 1.9)