Micropilot FMR50/FMR51/FMR52/FMR53/FMR54/FMR56/FMR57 FOUNDATION Fieldbus The "Expert" menu
Endress+Hauser 55
3.2.5 Expert → Sensor → Linearization
Overview
0%
100%
L
L
L
L
L
L
L
L2
L1
L‘1L‘2
L‘
L‘
L‘
L‘
L‘
L‘
L‘
M
M
M
M
M
H
h
h
M
0%
100%
100%
100%
0%
0%
100%
0%
0%
0%
100%
100%
B
A
C
D
E
F
G
H
2
1
L
L‘
d
d
A0019648
5 Linearization: Transformation of the level into a volume or weight; the transformation is dependent on
the shape of the vessel.
1 Selection of linearization type and unit
2 Parametrization of the linearization
A Linearization type = None
B Linearization type = Linear
C Linearization type = Table
D Linearization type = Pyramid bottom
E Linearization type = Conical bottom
F Linearization type = Angled bottom
G Linearization type = Horizontal cylinder
H Linearization type = Sphere
L Level before linearization (measured in level unit)
L' Level after linearization (corresponds to volume or weight)
M Maximum value
d Diameter
h Intermediate height