The "Expert" menu Micropilot FMR50/FMR51/FMR52/FMR53/FMR54/FMR56/FMR57 FOUNDATION Fieldbus
42 Endress+Hauser
Options None
Weak signal
Many obstacles
Small tanks (< 1m/3ft)
Changing DC values
Foam (> 5cm/0.16ft)
Factory setting None
Additional information Weak signal
Increases the sensitivity of the device in case of weak signals.
Many obstacles
Can improve the behavior of the device in case of many obstacles in a silo if strongly
damping media (e.g. flour) are measured.
Small tanks (< 1m/3ft)
Can improve the near-range measurement in small vessels.
Changing DC values
Adjusts the signal evaluation of the device to strongly changing DC values of the
measured medium.
Foam (> 5cm/0.16ft)
Adjusts the signal evaluation of the device to conditions with foam > 5 cm (1.6 in).
Application parameter (1126)
Navigation Expert → Sensor → Application parameter
Write access Read-only parameter
Description Indicates whether parameters have been changed after the basic setup.
Display options Changed
Not changed
Additional information Indicates whether a setting which depends on the application parameters (e.g. "Process
property", Advanced process conditions", "Tank type" and "Pipe diameter") has been changed
after the basic setup.
Changed
Parameters have been changed. The device is no longer in the state defined by the
application parameters.
Not changed
There have been no changes. The device is still in the state defined by the application
parameters.