The "Expert" menu Micropilot FMR50/FMR51/FMR52/FMR53/FMR54/FMR56/FMR57 FOUNDATION Fieldbus
46 Endress+Hauser
Description of parameters
Distance offset (2309)
Navigation Expert → Sensor → Level → Distance offset
Write access Maintenance
Description Definition of the distance offset
Input range -200.000,0 ... +200.000,0 mm (-60 ... +60 ft)
Factory setting 0.0 mm
Additonal Information The value speicified in this parameter is added to the measured distance between the
reference point of the measurement and the level echo.
Positive values increase the distance and thus decrease the level.
Negative values decrease the distance and thus increase the level.
D0
D
D0
R R
DD(<0)
DD(>0)
D
A0016081
 2 Effect of the "Distance offset" parameter
ΔD Distance offset
D0 Measured distance (as indicated in "Expert → Sensor → Distance → Distance")
D Corrected distance (is used to calculate the level)
R Reference point
A value entered into this parameter is taken into account in the calculation of the
level. However, it does not influence the indicated distance.
Distance (2231)
Navigation Expert → Sensor → Level → Distance
Write access Read-only parameter