EasyManua.ls Logo

Endress+Hauser Micropilot FMR56 - Connecting the Measuring Device

Endress+Hauser Micropilot FMR56
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR56, FMR57 FOUNDATION Fieldbus Electrical connection
Endress+Hauser 41
7.2 Connecting the measuring device
L
WARNING
Risk of explosion!
Observe applicable national standards.
Comply with the specifications in the Safety Instructions (XA).
Use specified cable glands only.
Check to ensure that the power supply matches the information on the nameplate.
Switch off the power supply before connecting the device.
Connect the potential matching line to the outer ground terminal before applying the
power supply.
Required tools/accessories:
For devices with a cover lock: Allen key AF3
Wire stripper
When using stranded cables: One ferrule for every wire to be connected.
7.2.1 Opening connection compartment cover
2.
1.
3 mm
A0021490
1. Loosen the screw of the securing clamp of the connection compartment cover using an
Allen key (3 mm) and turn the clamp 90 ° clockwise.
2. Afterwards unscrew connection compartment cover and check lid gasket, replace if
necessary.