EasyManua.ls Logo

Endress+Hauser Micropilot FMR56 - 9 Commissioning Via Wizard

Endress+Hauser Micropilot FMR56
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR56, FMR57 FOUNDATION Fieldbus Commissioning via wizard
Endress+Hauser 51
9 Commissioning via wizard
A wizard guiding the user through the initial setup is available in FieldCare and DeviceCare
2)
.
1. Connect the device to FieldCare or DeviceCare (for details refer to the "Operating
options" chapter of the Operating Instructions).
2. Open the device in FieldCare or DeviceCare.
The dashboard (home page) of the device appears:
1
A0027720
1 "Commissioning" button calls up the wizard.
3. Click on "Commissioning" to call up the wizard.
4. Enter or select the appropriate value for each parameter. These values are immediately
written to the device.
5. Click "Next" to switch to the next page.
6. After finishing the last page, click "End of sequence" to close the wizard.
If the wizard is cancelled before all necessary parameters have been set, the device may
be in an undefined state. A reset to the default settings is recommended in this case.
2) DeviceCare is available for download at www.software-products.endress.com. The download requires a
registration in the Endress+Hauser software portal.