Levelflex M 3 Function group "basic setup" (00)
Endress+Hauser 17
FMP43:
FMP41C, FMP45:
The lower group applies to very loose or loosened bulk solids.
Reduction of the max. possible measuring range by means of:
• extremely loose surfaces of bulk solids, e.g. bulk solids with low piled density when filled pneu-
matically.
• Build-up, primarily of moist products.
!
Note!
Due to the high diffusion rate of ammonia it is recommended to use the FMP45 with gas-tight bush-
ing for measurements in this medium.
Media group DC (εr) Typical liquids Typ. measuring range
1 1.4 to 1.6 Condensed gases, e.g. N
2
, CO
2
—
2 1.6 to 1.9
Liquefied gas, e.g. propane
Solvent
Frigen / Freon
Palm oil
4 m (354")
3 1.9 to 2.5 Mineral oils, fuels
4 2.5 to 4
Benzene, styrene, toluene
Furan
Naphthalene
5 4 to 7
Chlorobenzene, chloroform
Cellulose spray
Isocyanate, aniline
6 >7
Aqueous solutions
Alcohols
Acids, alkalis
Media group DC (
εr) Typical liquids Typ. measuring range
1 1.4...1.6 Condensed gases, e.g. N
2
, CO
2
4 m (157"), when installed in metallic pipes
2 1.6...1.9
Liquefied gas, e.g. Propane
Solvent
Frigen / Freon
Palm oil
9 m (354")
3 1.9...2.5 Mineral oils, fuels 12 m (472")
4 2.5...4
Benzene, styrene, toluene
Furan
Naphthalene
16 m (629")
5 4...7
Chlorobenzene, chloroform
Cellulose spray
Isocyanate, aniline
25 m (984")
6 >7
Aqueous solutions
Alcohols
Acids, alkalis
30 m (1181")