11 Function group "diagnostics" (0A) Levelflex M
64 Endress+Hauser
11.4 Function "reset" (0A3)
"
Caution!
A reset sets the instrument back to the factory settings. This can lead to an impairment of the meas-
urement. Generally, you should perform a basic setup again following a reset.
A reset is only necessary:
if the instrument no longer functions
if the instrument must be moved from one measuring point to another
if the instrument is being de-installed /put into storage/installed
Entry ("reset" (0A3)):
333 = customer parameters (HART)
33333 = customer parameters (PROFIBUS PA and FOUNDATION Fieldbus)
333 = reset customer parameters for HART
33333 = reset customer parameters for PROFIBUS PA and FOUNDATION Fieldbus
This reset is recommended whenever an instrument with an unknown 'history' is to be used in an
application:
The Levelflex is reset to the default values.
The customer specific tank map is not deleted.
A linearisation is switched to "linear" although the table values are retained. The table can be
reactivated in the "linearisation" (04) function group.
List of functions that are affected by a reset:
The mapping can also be deleted in the "cust. tank map" (055) function of the "extended
calibr." (05) function group.
A complete “basic setup" (00) must be activated.
ENDRESS + HAUSER
E
+
tank properties (002)
•medium cond. (003)
process proper. (004)
empty calibr. (005)
full calibr. (006)
output on alarm (010)
output on alarm (011)
outp. echo loss (012)
ramp %span/min (013)
delay time (014)
safety distance (015)
in safety dist. (016)
overspill protection (018)
end of probe (030)
level/ullage (040)
linearisation (041)
customer unit (042)
max. scale (046)
diameter vessel (047)
check distance (051)
range of mapping (052)
start mapping (053)
offset (057)
output damping (058)
low output limit (062)
curr. output mode (063)
fixed cur. value (064)
4mA value (068)
language (092)
back to home (093)
format display (094)
no of decimals (095)
sep. character (096)
unlock parameter (0A4)