Loading...
9 Function group "envelope curve" (0E) Levelflex M
58 Endress+Hauser
Horizontal-Zoom-Modus
Press
O or S, to switch to the envelope curve navigation. You are then in Horizontal Zoom mode.
Either or is displayed.
You now have the following options:
O increases the horizontal scale.
S decreases the horizontal scale.
Move-Modus
Then press
F, to switch to Move mode. Either or is displayed.
You now have the following options:
O shifts the curve to the right.
S shifts the curve to the left.
Vertical-Zoom-Modus
Press
F, once more to switch to Vertical Zoom mode is displayed.
You now have the following options:
O increases the vertical scale.
S decreases the vertical scale.
The display icon shows the current zoom factor ( to ).
Exiting the navigation
•Press
F again to run through the different modes of the envelope curve navigation.
Press
O and S to exit the navigation. The set increases and shifts are retained. Only when you
reactivate the "recording curve" (0E2) function does the Levelflex use the standard display again.
O
S
O
S
O
S
After 3 s, the following message appears
ENDRESS + HAUSER
E
+

Table of Contents