Design Proline Promag 300 HART
18 Endress+Hauser
NOTICE
When the SIL locking code is entered, the device parameters that affect the safety-
related output signal are locked and write-protected. It is still possible to read the
parameters.
When SIL locking is enabled, restrictions apply on all communication options, such as the
service interface (CDI-RJ45), HART protocol and WirelessHART protocol, local display and
WLAN.
Deactivation of the SIL mode →  28.