Appendix Proline Promag 300 HART
44 Endress+Hauser
• Sensor electronics module (ISEM)
• I/O modules
• Terminals
• Connection compartment cover
• Electronics compartment cover
• Seal sets for electronics compartment cover
• Securing clamps for electronics compartment cover
• Cable glands
Installation Instructions: see the Download Area under www.endress.com.
The replaced component must be sent to Endress+Hauser for the purpose of fault analysis if
the device has been operated in a protective system and a device error cannot be ruled out. In
this case, always enclose the "Declaration of Hazardous Material and Decontamination" with
the note "Used as SIL device in protection system" when returning the defective device. Please
refer to the "Return" section in the Operating Instructions .
8.3 Modification
Modifications are changes to devices with SIL capability already delivered or installed.
‣
Modifications to devices with SIL capability are usually performed in the Endress+Hauser
manufacturing center.
‣
Modifications to devices with SIL capability onsite at the user's plant are possible following
approval by the Endress+Hauser manufacturing center. In this case, the modifications must
be performed and documented by an Endress+Hauser service technician.
‣
Modifications to devices with SIL capability by the user are not permitted.
8.4 Taking out of service
The requirements of IEC 61508-1:2010 Section 7.17 must be observed when taking the
device out of service.
8.5 Disposal
For detailed information on the device disposal, see the Operating Instructions
9 Appendix
9.1 Structure of the measuring system
9.1.1 System components
An example of the devices in the measuring system is shown in the following graphic.