Proline Promag 300 HART Proof testing
Endress+Hauser 35
7.1.4 Connecting the test
1. Re-activate the locked SIL mode →  21.
2. Deactivate bypassing of safety function of process control system.
3. Document results of proof test in accordance with the safety management guidelines
applicable to the system.
NOTICE
At least 99 % of dangerous, undetected failures can be detected using the test sequences
described (PTC = 0.99). The influence of systematic errors on the safety function is not
fully covered by the test. Systematic errors can be caused, for example, by medium
properties, operating conditions or buildup.
If one of the test criteria from the test sequences described above is not fulfilled, the device
may no longer be used as part of a protective system.
Take measures to reduce systematic errors. For detailed information on the orientation,
medium properties and operating conditions, see the "Operating Instructions" →  10
7.2 Test sequence C (PTC = 98 %)
Testing with a secondary standard (volume flow)
Inspection and on-site visual check
Preparation
Byassing of safety function of process control system, to prevent accidental activation of the
safety function.
Deactivate the locked SIL mode →  28.
7.2.1 Test sequence: Testing with a secondary standard (volume flow)
Testing with a secondary standard (volume flow) Verification of the measured value for liquid
and gaseous volume flow by comparing with a secondary standard.
The measured values (3 to 5 measuring points) are checked with a secondary standard on an
installed device (mobile calibration rig or calibrated reference device) or on a factory
calibration rig following device removal.
The measured values of the secondary standard and the device under test (DUT) are compared
using one of the following methods:
Comparison by reading off the digital measured value
Compare the digital measured value of the secondary standard against the measured value
display of the DUT at the logic subsystem (process control system or safety-related PLC).
Comparison of the measured value by measuring the current
Requirements for the measuring equipment:
DC current measuring uncertainty ±0.2 %
DC current resolution 10 µA
1. Measure the current at the DUT using an external, traceably-calibrated ammeter.