Proline Promag H 300
Endress+Hauser 9
Flow characteristic values in SI units
Nominal
diameter
Recommended
flow
Factory settings
min./max. full scale value
(v ~ 0.3/10 m/s)
Current output full scale
value
1)
(v ~ 2.5 m/s)
Pulse value
1)
(~ 2 pulse/s)
Low flow cut off
(v ~ 0.04 m/s)
[mm] [in] [dm
3
/min] [dm
3
/min] [dm
3
] [dm
3
/min]
2 1/12 0.06 to 1.8 0.5 0.005 0.01
4 1/8 0.25 to 7 2 0.025 0.05
8 3/8 1 to 30 8 0.1 0.1
15 ½ 4 to 100 25 0.2 0.5
25 1 9 to 300 75 0.5 1
40 1 ½ 25 to 700 200 1.5 3
50 2 35 to 1 100 300 2.5 5
65 – 60 to 2 000 500 5 8
80 3 90 to 3 000 750 5 12
100 4 145 to 4 700 1200 10 20
125 5 220 to 7 500 1850 15 30
150 6 20 to 600 m
3
/h 150 m
3
/h 0.03 m
3
2.5 m
3
/h
1) HART only
Flow characteristic values in US units
Nominal
diameter
Recommended
flow
Factory settings
min./max. full scale value
(v ~ 0.3/10 m/s)
Current output full scale
value
1)
(v ~ 2.5 m/s)
Pulse value
1)
(~ 2 pulse/s)
Low flow cut off
(v ~ 0.04 m/s)
[in] [mm] [gal/min] [gal/min] [gal] [gal/min]
1/12 2 0.015 to 0.5 0.1 0.001 0.002
1/8 4 0.07 to 2 0.5 0.005 0.008
3/8 8 0.25 to 8 2 0.02 0.025
½ 15 1 to 27 6 0.05 0.1
1 25 2.5 to 80 18 0.2 0.25
1 ½ 40 7 to 190 50 0.5 0.75
2 50 10 to 300 75 0.5 1.25
3 80 24 to 800 200 2 2.5
4 100 40 to 1 250 300 2 4
5 125 60 to 1 950 450 5 7
6 150 90 to 2 650 600 5 12
1) HART only
To calculate the measuring range, use the Applicator sizing tool → 93
Recommended measuring range
"Flow limit" section → 48