Proline Promag P 200
10 Endress+Hauser
Intrinsically safe values
Type of protection Ex ia
Order code for
"Output"
Output type Intrinsically safe values
Option A 4-20 mA HART U
i
= DC 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0 mH
C
i
= 5 nF
Option B 4-20 mA HART U
i
= DC 30 V
I
i
= 300 mA
P
i
=1 W
L
i
= 0 mH
C
i
= 5 nF
Pulse/frequency/switch output U
i
= DC 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0 mH
C
i
= 6 nF
IS type of protection
Order code for
"Output"
Output type Intrinsically safe values
Option A 4-20 mA HART U
i
= DC 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0 mH
C
i
= 5 nF
Option B 4-20 mA HART U
i
= DC 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0 mH
C
i
= 5 nF
Pulse/frequency/switch output U
i
= DC 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0 mH
C
i
= 6 nF
Low flow cut off
The switch points for low flow cut off are user-selectable.
Galvanic isolation
All outputs are galvanically isolated from one another.
Protocol-specific data HART
Manufacturer ID
0x11
Device type ID 0x48
HART protocol revision 6.0
Device description files (DTM,
DD)
Information and files under:
www.endress.com

Table of Contents

Related product manuals