Proline Promag P 200
Endress+Hauser 25
Pressure tightness
Liner: PFA
Nominal diameter Limit values for absolute pressure in [mbar] ([psi]) for fluid temperatures:
[mm] [in] +25 °C (+77 °F) +80 °C (+176 °F) +100 to +180 °C (+212 to
+356 °F)
25 1 0 (0) 0 (0) 0 (0)
32 – 0 (0) 0 (0) 0 (0)
40 1 ½ 0 (0) 0 (0) 0 (0)
50 2 0 (0) 0 (0) 0 (0)
65 – 0 (0) – (–) 0 (0)
80 3 0 (0) – (–) 0 (0)
100 4 0 (0) – (–) 0 (0)
125 – 0 (0) – (–) 0 (0)
150 6 0 (0) – (–) 0 (0)
200 8 0 (0) – (–) 0 (0)
Liner: PTFE
Nominal diameter
Limit values for absolute pressure in [mbar] ([psi]) for fluid temperatures:
[mm] [in] +25 °C (+77 °F) +80 °C (+176 °F) +100 °C (+212 °F) +130 °C (+266 °F)
15 ½ 0 (0) 0 (0) 0 (0) 100 (1.45)
25 1 0 (0) 0 (0) 0 (0) 100 (1.45)
32 – 0 (0) 0 (0) 0 (0) 100 (1.45)
40 1 ½ 0 (0) 0 (0) 0 (0) 100 (1.45)
50 2 0 (0) 0 (0) 0 (0) 100 (1.45)
65 – 0 (0) – (–) 40 (0.58) 130 (1.89)
80 3 0 (0) – (–) 40 (0.58) 130 (1.89)
100 4 0 (0) – (–) 135 (1.96) 170 (2.47)
125 – 135 (1.96) – (–) 240 (3.48) 385 (5.58)
150 6 135 (1.96) – (–) 240 (3.48) 385 (5.58)
200 8 200 (2.90) – (–) 290 (4.21) 410 (5.95)
Flow limit
The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum velocity
of flow is between 2 to 3 m/s (6.56 to 9.84 ft/s). Also match the velocity of flow (v) to the physical properties
of the fluid:
•
v < 2 m/s (6.56 ft/s): for abrasive fluids (e.g. potter's clay, lime milk, ore slurry)
•
v > 2 m/s (6.56 ft/s): for fluids producing buildup (e.g. wastewater sludges)
For an overview of the measuring range full scale values, see the "Measuring range" section (® ä 5)
Pressure loss
• No pressure loss occurs if the sensor is installed in a pipe with the same nominal diameter.
•
Pressure losses for configurations incorporating adapters according to DIN EN 545 (® ä 20)
System pressure
• Never install the sensor on the pump suction side in order to avoid the risk of low pressure, and thus damage
to the liner.
• Furthermore, install pulse dampers if reciprocating, diaphragm or peristaltic pumps are used.
•
For information on the liner's resistance to partial vacuum (® ä 25)
•
For information on the measuring system's resistance to vibration and shock (® ä 22), (® ä 22)