Measuring system
The device consists of a transmitter and a sensor.
One device version is available: compact version, transmitter and sensor form a mechanical unit.
Transmitter
Promag 200
Materials:
Aluminum coating AlSi10Mg
Configuration:
External operation via four-line, illuminated local display with touch control
and guided menus ("Make-it-run" wizards) for applications
Via operating tools (e.g. FieldCare)
A0013471
Sensor
Promag P
Nominal diameter range: DN 15 to 200 (½ to 8")
Materials:
Sensor housing: aluminum coated AlSi10Mg
Measuring tubes: stainless steel 1.4301/304, 1.4306/304L
Liner: PFA, PTFE
Process connections: 1.0425/316L, 1.4571/316L, A105, C22, F316L, FE
410W B, HII, S235JRG2
Electrodes: 1.4435/316L, Alloy C-22, tantalum, platinum
Seals: as per DIN EN 1514-1
Ground disks: 1.4435/316L, Alloy C-22, tantalum
A0017703
Input
Measured variable Direct measured variables
Volume flow (proportional to induced voltage)
Calculated measured variables
Mass flow
Measuring range
Typically v = 0.01 to 10 m/s (0.03 to 33 ft/s) with the specified accuracy
Flow characteristic values in SI units
Nominal
diameter
Recommended
flow
Factory settings
min./max. full scale value
(v ~ 0.310 m/s/
Full scale value current
output
(v ~ 2.5 m/s)
Pulse value
(~ 2 pulse/s)
Low flow cut off
(v ~ 0.04 m/s)
[mm] [in] [dm
3
/min] [dm
3
/min] [dm
3
] [dm
3
/min]
15 ½ 4 to 100 25 0.2 0.5
25 1 9 to 300 75 0.5 1
32 15 to 500 125 1 2
40 1 ½ 25 to 700 200 1.5 3
50 2 35 to 1 100 300 2.5 5
65 60 to 2 000 500 5 8
80 3 90 to 3 000 750 5 12
100 4 145 to 4 700 1 200 10 20
125 220 to 7 500 1 850 15 30