Proline Promag P 200
18 Endress+Hauser
2.5
[%]
2.0
1.5
1.0
0.5
0
0.5 %
0
1
2 4 6 8 10 [m/s]
v
5 10 15 20 25 30 32 [ft/s]0
A0003200
å 10 Maximum measured error in % o.r.
Repeatability
o.r. = of reading
max. ±0.2 % o.r. ±2 mm/s (0.08 in/s)
Influence of ambient
temperature
o.r. = of reading; o.f.s. = of full scale value
Current output
Additional error, in relation to the span of 16 mA:
Temperature coefficient at
zero point (4 mA)
0.02 %/10 K, max. 0.35 % over the entire temperature range
–40 to +60 °C (–40 to +140 °F)
Temperature coefficient
with span (20 mA)
0.05 %/10 K, max. 0.5 % over the entire temperature range
–40 to +60 °C (–40 to +140 °F)
Pulse/frequency output
Temperature coefficient
Max. ±100 ppm o.r.
Installation
No special measures such as supports are necessary. External forces are absorbed by the construction of the
device.
Mounting location
Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the next pipe elbow: h =
³ 2 × DN
h
A0017061