10 Group PULSE/FREQUENCY OUTPUT Device Functions Proline Promass 80
40 Endress+Hauser
VALUE SIMULATION
PULSE
! Note!
This function is not available unless the COUNTDOWN option was selected in the
SIMULATION PULSE function.
Use this function to specify the number of pulses (e.g. 50) which are output during the
simulation. This value is used to test downstream devices and the measuring device itself.
The pulses are output with the pulse width specified in the PULSE WIDTH function. The
on/off ratio is 1:1.
Simulation is started once the specified value is confirmed with the F key. The display
remains at 0 if the specified pulses have been output.
User input:
0...10 000
Factory setting:
0
! Note!
Simulation is started by confirming the simulation value with the F key. The simulation
can be switched off again via the SIMULATION PULSE function.
" Caution!
The setting is not saved if the power supply fails.
Function description PULSE/FREQUENCY OUTPUT