Device Functions Proline Promass 80 17 Group SUPERVISION
Endress+Hauser 59
17 Group SUPERVISION
Function description SUPERVISION
CURRENT SYSTEM
CONDITION
Use this function to check the current system status.
Display:
“SYSTEM OK” or the fault / notice message with the highest priority.
PREVIOUS SYSTEM
CONDITIONS
Use this function to view the fifteen most recent fault and notice messages since
measuring last started.
Display:
The 15 most recent fault or notice messages.
ASSIGN SYSTEM ERROR Use this function to view all system messages and the associated error categories (fault
message or notice message). If you select a single system fault you can change its error
category.
Display:
List of system errors
! Note!
Press the F key twice to call the ERROR CATEGORY function.
•Use the OS key combination or select CANCEL in the system error list to exit the
function.
ERROR CATEGORY Use this function to define whether a system fault triggers a notice message or a fault
message. If you select “FAULT MESSAGES”, all outputs respond to a fault in accordance
with their defined error response patterns.
Options:
NOTICE MESSAGES (display only)
FAULT MESSAGES (outputs and display)
! Note!
Press the F key twice to call the
function ASSIGN SYSTEM ERROR.
•Use the OS key combination to exit the function.
ASSIGN PROCESS ERROR Use this function to view all process errors and the associated error categories (fault
message or notice message). If you select a single process error you can change its error
category.
Display:
List of process errors
! Note!
Press the F key twice to call the ERROR CATEGORY function.
•Use the OS key combination or select CANCEL in the process error list to exit the
function.