12 Group STATUS INPUT Device Functions Proline Promass 80
46 Endress+Hauser
12 Group STATUS INPUT
Function description STATUS INPUT
This group is not available unless the measuring device is fitted with a status input I/O module.
ASSIGN STATUS INPUT Use this function to assign a switching function to the status input.
Options:
OFF
RESET TOTALIZER 1
POSITIVE ZERO RETURN
ZEROPOINT ADJUSTMENT
RESET TOTALIZER 2
RESET ALL TOTALIZERS
Factory setting:
OFF
! Note!
Positive zero return is active as long as the active level is available at the status input
(continuous signal). All other assignments react to a change in level (pulse) at the status
input.
ACTIVE LEVEL Use this function to define whether the assigned function (see ASSIGN STATUS INPUT
function) is released when the signal level is present (HIGH) or not present (LOW).
Options:
HIGH
LOW
Factory setting:
HIGH
MIN. PULSE WIDTH Use this function to define a minimum width which the input pulse must achieve in
order to trigger the defined switching function.
User input:
20...100 ms
Factory setting:
50 ms
SIMULATION STATUS
INPUT
Use this function to activate simulation of the status input,
in other words to trigger the function assigned to the status input (see the ASSIGN
STATUS INPUT function on Page 46).
Options:
OFF
ON
Factory setting:
OFF
! Note!
• The “SIMULATION STATUS INPUT” message indicates that simulation is active.
• The measuring device continues to measure while simulation is in progress, i.e. the
current measured values are output correctly via the outputs.
" Caution!
The setting is not saved if the power supply fails.