Device Functions Proline Promass 80 12 Group STATUS INPUT
Endress+Hauser 47
VALUE SIMULATION
STATUS INPUT
! Note!
This function is not available unless the ON setting was selected in the SIMULATION
STATUS INPUT function.
Use this function to select the level to be assumed at the status input during the
simulation.
Options:
HIGH
LOW
Factory setting:
LOW
" Caution!
The setting is not saved if the power supply fails.
Function description STATUS INPUT