Device Functions Proline Promass 80 10 Group PULSE/FREQUENCY OUTPUT
Endress+Hauser 29
10 Group PULSE/FREQUENCY OUTPUT
Function description PULSE/FREQUENCY OUTPUT
OPERATION MODE Use this function to configure the output as a pulse or frequency output. The functions
available in this function group vary, depending on which option you select here.
Options:
PULSE
FREQUENCY
Factory setting:
PULSE
ASSIGN FREQUENCY
! Note!
This function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function.
Use this function to assign a measured variable to the frequency output.
Options:
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
DENSITY
REFERENCE DENSITY
TEMPERATURE
Factory setting:
MASS FLOW
! Note!
If you select OFF, the only functions shown in this function group are ASSIGN
FREQUENCY and OPERATION MODE.
END VALUE FREQ.
! Note!
This function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function.
Use this function to define a full scale frequency for the frequency output. You define the
associated measured value of the measuring range in the VALUE F HIGH function
described on Page 30.
User input:
4-digit fixed-point number: 2...1000 Hz
Factory setting:
1000 Hz
Example:
• VALUE F HIGH = 1000 kg/h, full scale frequency = 1000 Hz: i.e. a frequency of 1000
Hz is output at a flow of 1000 kg/h.
• VALUE F HIGH = 3600 kg/h, full scale frequency = 1000 Hz: i.e. a frequency of 1000
Hz is output at a flow of 3600 kg/h.
! Note!
In the FREQUENCY operating mode the output signal is symmetrical (on/off ratio = 1:1).
At low frequencies the pulse duration is limited to a maximum of 2 seconds, i.e. the
on/off ratio is no longer symmetrical.