Device Functions Proline Promass 80 18 Group SIMULATION SYSTEM
Endress+Hauser 61
18 Group SIMULATION SYSTEM
Function description SIMULATION SYSTEM
SIMULATION
FAILSAFE MODE
Use this function to set all inputs, outputs and totalizer to their defined fault response
modes, in order to check whether they respond correctly. During this time, the words
“SIMULATION FAILSAFE MODE” appear on the display.
Options:
OFF
ON
Factory setting:
OFF
SIMULATION
MEASURAND
Use this function to set all inputs, outputs and totalizer to their defined flow-response
modes, in order to check whether they respond correctly. During this time, the words
“SIMULATION MEASURAND” appear on the display.
Options:
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
DENSITY
REFERENCE DENSITY
TEMPERATURE
Factory setting:
OFF
" Caution!
• The measuring device cannot be used for measuring while this simulation is in
progress.
• The setting is not saved if the power supply fails.
VALUE SIMULATION
MEASURAND
! Note!
This function is not available unless the SIMULATION MEASURAND function is active.
Use this function to define a selectable value (e.g. 12 kg/s). This value is used to test
downstream devices and the measuring device itself.
User input:
5-digit floating-point number
Factory setting:
0 kg/h (MASS FLOW)
0 m
3
/h (VOLUME FLOW)
0 Nm
3
/h(CORRECTED VOLUME FLOW)
0 kg/l(DENSITY)
0 kg/NI(REFERENCE DENSITY)
0 °C(TEMPERATURE)
" Caution!
The setting is not saved if the power supply fails.