3 Group SYSTEM UNITS Device Functions Proline Promass 80
8 Endress+Hauser
3 Group SYSTEM UNITS
Function description SYSTEM UNITS
You can select the unit for the measured variable in this function group.
UNIT MASS FLOW Use this function to select the unit for displaying the mass flow (mass/time).
The unit you select here is also valid for:
Current output
Frequency output
Switching points (limit value for mass flow, flow direction)
Low flow cut off
Options:
Metric:
gram
g/s; g/min; g/h; g/day
kilogram
kg/s; kg/min; kg/h; kg/day
Metric ton
t/s; t/min; t/h; t/day
US:
ounce
oz/s; oz/min; oz/h; oz/day
pound
lb/s; lb/min; lb/h; lb/day
ton
ton/s; ton/min; ton/h; ton/day
Factory setting:
Country dependent (kg/h or US lb/day)
UNIT MASS Use this function to select the unit for displaying the mass.
The unit you select here is also valid for:
Pulse value (e.g. kg/p)
Options:
Metric
g; kg; t
US
oz; lb; ton
Factory setting:
Country dependent (kg or US lb)
! Note!
The unit for the totalizer is independent of your choice here, it is selected separately in
the TOTALIZER 1/2 function group (see Page 20).