Device Functions Proline t-mass 65 5 Group QUICK SETUP
Endress+Hauser 13
5 Group QUICK SETUP
Function description QUICK SETUP
QUICK SETUP
COMMISSION
Use this function to start the Setup menu for commissioning.
Options:
NO
YES
Factory setting:
NO
! Note!
You will find a flowchart of the Quick Setup Commissioning menu on Page 14. For more
information on Setup menus, please refer to the Operating Instructions Proline t-mass 65,
BA111D/06/en/…
T-DAT
SAVE/LOAD
Use this function to save the parameter settings / configuration of the transmitter in a
transmitter DAT (HistoROM/T-DAT), or to load the parameter settings from the
HistoROM/T-DAT into the EEPROM (manual safety function).
Application examples:
• After commissioning, the current measuring point parameters can be saved to the
HistoROM/T-DAT as a backup.
• If the transmitter is replaced for some reason, the data from the HistoROM/T-DAT can
be loaded into the new transmitter (EEPROM).
Options:
CANCEL
SAVE (from EEPROM to HistoROM/T-DAT)
LOAD (from the HistoROM/T-DAT into EEPROM)
Factory setting:
CANCEL
! Note!
• If the target device has an older software version, the message “TRANSM. SW-DAT” is
displayed during start-up. Then only the SAVE option is available.
•LOAD
This option is only possible if the target device has the same software version as, or a
more recent software version than, the source device.
•SAVE
This option is always available.