13 Group STATUS INPUT Device Functions Proline t-mass 65
48 Endress+Hauser
13 Group STATUS INPUT
Function description STATUS INPUT
! Note!
This group is not available unless the transmitter is fitted with a status input I/O module.
ASSIGN STATUS INPUT Use this function to assign a switching function to the status input.
Options:
OFF
RESET TOTALIZER 1
POSITIVE ZERO RETURN
ZERO POINT ADJUST
RESET TOTALIZER 2
RESET ALL TOTALIZERS
Factory setting:
OFF
! Note!
• Positive zero return is active as long as the active level is available at the status input
(continuous signal). All other assignments react to a change in level (pulse) at the sta-
tus input.
• If you select OFF, the only function shown in the STATUS INPUT group is this func-
tion ASSIGN STATUS INPUT.
ACTIVE LEVEL Use this function to define whether the assigned function (see ASSIGN STATUS INPUT
function) is released when the signal level is present (HIGH) or not present (LOW).
Options:
HIGH
LOW
Factory setting:
HIGH
MINIMUM
PULSE WIDTH
Use this function to define a minimum width which the input pulse must achieve in
order to trigger the defined switching function.
User input:
20 to 100 ms
Factory setting:
50 ms
ACTUAL STATUS INPUT Use this function to define whether the assigned function (see ASSIGN STATUS INPUT
function) is released when the signal level is present (HIGH) or not present (LOW).
Options:
HIGH
LOW
Factory setting:
HIGH